Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-32 Planmeca Proline CC panoramic x-ray
Technical Manual
Leave the ball phantom tool in position and place alignment tool adapter to the sensor head.
Place one end of the alignment ruler in the first primary slot (the largest one) and the other end to
the sensor alignment tool.
Figure
57
Loosen the four screws that hold the patient position mechanism in place and adjust its position
to the left or right until the mechanism is centered.
Figure
58
Make sure that the y-line on the ball phantom is parallel to the line on the underside of the align-
ment ruler. Note that the line on the ball phantom and the line on the alignment ruler do not have
to coincide, but they must be parallel, the deviation may be ±1mm.
Figure
59
Tighten the screws, and remove the alignment pin.
Take another ball phantom picture to check the alignment again. Readjust if necessary.
START
END
PLANMECA
No.50971
X-LINE
Y-LINE
Dimax2_adj1.eps
Alignment ruler
Screws
START
END
PLANMECA
No.50971
X-LINE
Y-LINE
Y
-lineca.EPS
Alignment ruler
Lines do not have to coincide
but they must be parallel
Ball phantom
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......