Planmeca Proline CC panoramic x-ray H-1
Chapter
Technical Manual
H
PARTS REPLACEMENT & REPAIR
1
COVERS
CAUTION
The unit contains hazardous voltages. Always disconnect the unit from mains by remov-
ing the plug from the wall outlet, and wait for 2 minutes before starting to remove covers.
1.1
Tube head covers
Remove the front cover of the tube head by unscrewing the 4 fixing screws at the back side.
Remove the back cover of the tube head by unscrewing the 2 fixing screws that can be seen
after removing the front cover. (Fig. 1, 1)
1.2
Secondary slot cover
Remove the secondary slot cover by unscrewing the 4 screws at the cassette head (Fig. 1, 2).
1.3
Cassette carriage back cover
Remove the secondary slot cover.
Remove the back cover of the cassette head by unscrewing the two fixing screws behind the
cover of the secondary slot. (Fig. 1, 3)
Figure
1
THIS X-R
AY UNIT MAY BE DANGEROUS TO PATIENT
AND OPERATO
R UNLES
S SAFE EXPO
SURE FACTOR
S
AND O
PERATING INSTRUCTIO
NS ARE OBS
ERV
ED
ATTENTION CET APPAREIL A RAYONS
X PEUT ETR
E
DAN
GEREUX PO
UR LE
PATIENT ET L'OPER
ATEUR
SI LES F
ACTEURS DE SECURITE ET LE M
ODE
D'EM
PL
OI
NE
SONT PA
S SCRUPUL
EUSEMENT
RE
SPECTES
WARNING
TOTA
L FIL
TRA
TIO
N 2,
5 mm Al
FILT
RAT
ION TOTALE 2,5 mm equ. A
l
1
2
3
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......