Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-2 Planmeca Proline CC panoramic x-ray
Technical Manual
1
REQUIRED TOOLS
•
Alignment ruler (part number 50973). For checking the position of the patient positioning
mechanism.
•
Ball phantom (part number 50971). For checking the position of the patient positioning mech-
anism and the positioning lights.
•
Frankfort plane alignment tool (part number 50977). Used with the ball phantom for checking
the position of the Frankfort plane light.
Alignr
ul.eps
Alignment ruler
START
END
PLANMECA
No.50971
X-LINE
Y-L
IN
E
Ball phantom
START
END
PLANMECA
No.50971
X-LINE
Y-LINE
PLANMECA OY
NO.50977
Frankfort plane
alignment tool