Chapter
D - ADJUSTMENT & CALIBRATION
Planmeca Proline CC panoramic x-ray D-43
Technical Manual
9.6
Soft tissue filter adjustment
Check the radiation beam position according to instructions given in section “Checking the soft
tissue filter and the filter positioning light position” on page D-37. If the left side of the radiation
beam is too far left or right the radiation beam must be repositioned.
Radiation beam too far left or right
Check the place of the radiation beam from the scale on the beam alignment tool (Fig. 50, 1).
Screw the cephalostat arm adjusting screw according to the table (Fig. 50, 2, 3).
For example: if the radiation beam extends 3mm passed the “0mm line” (Fig. 50, 1), screw the
adjusting screw about 0.6 turn clockwise.
If the radiation beam does not reach the “0mm line”, the screw must be screwed counterclock-
wise according to the same table.
Figure
50
NOTE
Place the ceph head support alignment tools in the ear post holders and check the cepha-
lostat arm adjustment according to instruction given in section “Checking the head sup-
port position” on page D-34.
1
2
3
4
3
6
9
12
15
4
5
6
PLANMECA PM 2002 Proline CC Cephalostat 18x24
No.652232
Posiotioning light 25mm
Posiotioning light 0mm
Soft tissue filter 0mm
mm
9
6
3 0
3
6
9
12 mm
1
2
3
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......