Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline CC panoramic x-ray E-5
Technical Manual
Unscrew the two attachment screws of the quick connector mechanism cover with the 4mm allen
screw and slide the cover from its position.
NOTE
In case the x-ray is a pan/ceph unit with two fixed sensor heads, the cover is attached to
two attachment pins. Unscrew the two attachment screws with the 2.5mm allen key and
remove the cover by sliding it horizontally from its position.
Figure
2
Loosen the four screws located on the lock disc.
Figure
3
Cover
Lock disc