Chapter
D - ADJUSTMENT & CALIBRATION
Planmeca Proline CC panoramic x-ray D-39
Technical Manual
The radiation beam must appear within borders of the rectangle marked on the beam alignment
tool and to the right of the line marked “Soft tissue filter 0mm”.
Figure
43
If the left-hand side of the radiation beam does not reach the line marked “Soft tissue filter 0mm”
or extends passed the line the soft tissue filter is in the wrong position it must be repositioned.
Refer to the section “Soft tissue filter adjustment” on page D-43.
Move the soft tissue filter positioning light until it strikes the line on the beam alignment tool
marked “Positioning light 25mm”. The number 25 must appear on the on patient positioning dis-
play. If the display number is wrong it must be recalibrated. Refer to the section “Soft tissue filter
positioning light calibration and adjustment” on page D-40.
Figure
44
4
5
6
PLANMECA PM 2002 Proline Cephalostat 18x24 No. 652232
Soft tissue filter 0mm
Positioning light 0mm
Positioning light 25mm
“Soft tissue filter 0mm” line
Radiation beam
Area of soft tissue filter
4
5
6
PLANMECA PM 2002 Proline Cephalostat 18x24 No. 652232
Soft tissue filter 0mm
Positioning light 0mm
Positioning light 25mm
°
mm
“Positioning light 25mm” line
Light beam position
2 5
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......