Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline CC panoramic x-ray E-53
Technical Manual
Start the Dimax3 calibration program by double-clicking the Dimax2Tool.exe program icon
located in the Dimaxis program folder. The DimaxTool window appears.
Figure
90
Calibrating the sensor head
Select Proline from the Settings / Type pulldown menu.
Figure
91
The calibration must be performed with both 3x3 (enhanced resolution) and 4x4 (normal resolu-
tion) binning. Select the binning from the Settings / Binning pulldown menu.
Figure
92
Click the Dimax3 cephalometric exposure button. The pmpcal2 window with text Waiting
appears.
Figure
93
Press the ready key to drive the x-ray to the exposure position.
Dimax3 cephalometric
exposure
READY
PRET