Chapter
D - ADJUSTMENT & CALIBRATION
D-12 Planmeca Proline CC panoramic x-ray
Technical Manual
If the distance from the middle ball to the center of the tenth ball on the left-hand side is more
than 122mm and on the right-hand side less than 118mm then the patient positioning mechanism
must be moved away from the column (Fig. 16). Note, that the distance between the line on the
underside of the alignment ruler and the y-line on the ball phantom must be less than ±1mm, see
section “Adjustment in ball phantom’s y-direction” on page D-14.
Figure
16
If the distance from the middle ball to the center of the tenth ball on the left-hand side is less than
118mm and on the right-hand side more than 122mm then the patient positioning mechanism
must be moved towards the column (Fig. 17). Note, that the distance between the line on the
underside of the alignment ruler and the y-line on the ball phantom must be less than ±1mm, see
section “Adjustment in ball phantom’s y-direction” on page D-14.
Figure
17
If the distances from the middle ball to the center of the tenth ball on both the right and left are
more than 122mm then the patient positioning mechanism is too far backward and must be
moved forwards (Fig. 18).
Figure
18
ST
AR
T
END
PLA
NME
CA
No
.509
71
X-L
INE
Y-LINE
Right-hand side
Film
Less than 118mm
More than 122mm
STAR
T
END
PLA
NMEC
A
No.50971
X-LIN
E
Y-LINE
Right-hand side
Film
Less than 118mm
More than 122mm
START
END
PLANMEC
A
No.50971
X-LINE
Y-LINE
Film
More than 122mm
More than 122mm
Right-hand side
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......