Chapter
D - ADJUSTMENT & CALIBRATION
D-20 Planmeca Proline CC panoramic x-ray
Technical Manual
The midsagittal light beam should strike the black line on the front of the ball phantom.
Figure
30
If the light beams do not coincide with the black lines or they are not in focus, they should be
adjusted.
4.2
Positioning light adjustment
Bending the mirror bracket backwards or forwards will position the light beam.
Figure
31
Rotating the mirror bracket will adjust the light beam position in vertical direction.
Figure
32
START
END
PLANMECA
No.50971
X-LINE
Y-LINE
PLANMECA OY
NO.50977
Midsagittal light beam
Adj.beam r
ightleft.eps
Mirror bracket
Adj.beam updo
wn.eps