Chapter
G - TROUBLESHOOTING
Planmeca Proline CC panoramic x-ray G-11
Technical Manual
8
DISTURBANCES IN PRIMARY SLOT MECHANISM
8.1
Primary slot does not move
8.2
Primary slot does not stop to the right position
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Motor does not start
Keyboard out of order or working
improperly.
PCB out of order or connector
loose.
Motor out of order or connector
loose.
Motor starts but slot is not moving
Clutch between drawing screw and
step motor is loose.
Thread of drawing nut damaged.
Primary slot carriage stuck to guid-
ance wheel.
Drawing nut unscrewed from draw-
ing screw.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Slot No. 0 in the right position but
other slots in the wrong position.
Clutch glides.
Program error.
All slot positions are wrong.
Position of the tube head changed
during the transportation.
Slot has moved in relation to the
slot carriage.
Optical detector tumbler has
moved.
Optical detector PCB has moved.
Fixation of the drawing nut of the
slot carriage has moved.
Optical detector defective.
Slot positions not calibrated.
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......