Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-12 Planmeca Proline CC panoramic x-ray
Technical Manual
The beam image should appear within the borders of the rectangle marked on the alignment tool
(if the beam is not correctly aligned see section “Radiation beam adjustment” on page E-14).
Figure
15
If the beam from the first slot is correctly aligned you should check the beam alignment of the
other slots. Note that the height of the other beams will vary as the slots are of different sizes.
Press the horizontal segment selection key to select the next primary slot. The number 1 will
appear on the horizontal sector display. This is the second primary slot. Repeat the previously
described alignment checking procedure for this beam. If this beam is not aligned correctly refer
to section “Radiation beam adjustment” on page E-14 for information on how to align the beam.
When it has been checked the number 2 and number 3 slots must be checked.
Figure
16
If the x-ray unit is equipped with the Transversal slicing system, the slot number 1 is replaced
with the slot number 7. The radiation beam from slot number 7 must be slightly wider than the
NORMAL BEAM and the same height as the rectangle marked on the alignment tool.
Figure
17
UP
DOWN
PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN no. 50972
NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT
Primary slot number 0
Alignment rectangle
Beam
Beam alignment tool
1
UP
DOWN
PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN no. 50972
NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT
Primary slot number 1
Beam
Beam alignment tool
7
UP
DOWN
PM 2002 CC X-RAY BEAM ALIGNMENT SCREEN no. 50972
NOTE: DO NOT LEAVE SCREEN IN DAYLIGHT
Primary slot number 7
Beam
Beam alignment tool
Summary of Contents for Proline CC
Page 2: ......
Page 42: ...Chapter C HELP ERROR MESSAGES C 12 Planmeca Proline CC panoramic x ray Technical Manual...
Page 186: ...Chapter H PARTS REPLACEMENT REPAIR H 8 Planmeca Proline CC panoramic x ray Technical Manual...
Page 202: ...Chapter I FUNCTIONAL DESCRIPTION I 16 Planmeca Proline CC panoramic x ray Technical Manual...
Page 203: ...Planmeca Proline CC panoramic x ray J 1 Chapter Technical Manual J SCHEMATICS DIAGRAMS...
Page 204: ...Chapter J SCHEMATICS DIAGRAMS J 2 Planmeca Proline CC panoramic x ray Technical Manual...
Page 206: ......