Software Reference
86
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Status Window
Use the Status window to view the status of a prime, instrument
calibration, wash, or run while an operation is running or after the
operation is complete. To open the Status window, touch the
Status
.
Figure 26. Status Window
The Status window displays the following information:
Chip Status
- Displays the current status of the chip. The LabChip
GX Touch automatically detects the chip type from an embedded
RF ID tag in each chip.
•
Type
- Displays the type of chip in the instrument, or the assay
type selected for the chip if the chip supports multiple assays.
•
Chip Reagents
- Displays the time and number of samples
(whichever comes first) until the chip reagents expire. Reagent
expiration depends on the type of chip inserted in the
instrument. The expiration time and number of samples reset
whenever the front door is opened.
Prime/Calibrate Status
(Protein Clear HR and ProteinEXact chips
only) - Displays the status of the current or just completed prime
and calibration.