Preparing the Instrument to Run Samples
30
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Defining a New Plate Type (Continued)
•
The instrument will move to the specified number of wells in
each corner of the plate, punching holes in the tape.
•
The Punch test is acceptable if the punched hole is >0.5mm
from the edge of the well.
•
Holes may not be perfectly centered.
Possible sources of errors:
•
Sipper splay
•
Plate tolerances
•
Robot alignment error
•
If all holes are too close to the edge (<0.5mm) in the same
direction, adjust the X-Margin or Y-Margin values on the Add
Plate window.
8
Select the new plate name on the
when starting
the run.