Software Reference
95
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Diagnostics Tab
Use the Diagnostics tab to run the diagnostics tests on the LabChip
GX Touch or GXII Touch instrument. All tests should be run
periodically to verify proper operation of the instrument. To verify a
particular function, select specific tests to run. To open the
Diagnostics Tab, touch the
Tools
button on the
, and
then touch the
Diagnostics
button
Figure 32. Diagnostics Tab
The left side of the tab displays the tests to run on the instrument.
Expand each section to view all of the tests in each section. Tests
selected with a check mark will run. To skip a test, touch the check
box to clear the selection. Touch a section check box to select or
clear all of the tests in the section.
The right side of the tab displays the results of the tests. The icon
color indicates the status of each test:
•
Blue
- The test is in progress.
•
Yellow
- The test was skipped.
•
Green
- The test passed.
•
Red
- The test failed or was aborted.