Troubleshooting and Diagnostics
154
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Description of Instrument Diagnostic Tests
This section includes the following information about the diagnostic
tests: the test names, descriptions, potential failures, and the chip
required to run the test. Touch the
Limits Report
button to display
the limits for each diagnostic test.
•
Computer Resources Test on page 154
•
System Components Test on page 155
•
Chip Interface Test on page 155
•
•
Pressure Leak Test on page 156
•
Chip Temperature Test on page 156
•
Current Leakage Test on page 157
•
•
Plate Handler Test on page 157
•
•
HV Voltage Calibration Test on page 159
•
HV Current Calibration Test on page 159
•
(if the instrument is equipped with a
barcode reader)
Computer Resources Test
Any chip can be loaded
Description
Potential Cause of Failure
Memory Check
Available memory below 500 MB. System
may function with lower memory but
there is risk of failure if analyzing many
plates.
Disk Space Check
Available disk space below 4 GB. Risk of
losing data as disk space is used. Free
up space on local hard drive.