LabChip GX Touch/GXII Touch Instrument Description
127
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Touch
Screen
Displays the software to control the instrument. Touch the screen to
select the desired operations.
Power
Button
Turns the instrument and touch screen on. Illuminated when the
instrument is on. To turn the instrument off, select Start --> Shutdown
on the Windows desktop to shut down the software. Using the power
button to shut off the instrument may result in file corruption. Use only
in case of emergency or if the computer has locked up.
Status
Indicator
Lights
Indicate the status of the instrument:
Not Lit
- Power is off.
Solid Green
- Power is on, GX Touch software is active, but
instrument is idle.
Flashing Green
- Running an assay or a
Protein Clear HR/ProteinEXact calibration.
Solid Yellow
- Power is on, but GX Touch software is not running in
Windows.
Solid Red
- Power is on, GX Touch software is active, but it cannot
communicate with the instrument
Flashing Red
- The instrument has failed to perform an action issued
by the LabChip GX Touch software.
Front Door
Opens automatically to provide access to the chip. Use the LabChip
GX Touch software to open the Front Door. When the door opens, a
message displays reminding users to perform appropriate
maintenance procedures (see
).