Running an Assay
36
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Defining the Run Parameters (Continued)
21
To use the Sample Names file to select the wells to read, select
the
Use Sample Names File for Sample Selection
check box.
(For details, see
22
To use a file to supply the Expected Peaks:
a
Select the
Expected Peaks File
check box.
b
Touch the
Browse
button next to
Expected Peaks File
.
c
Select the name of the GEP file that contains the expected
peaks, and touch the
Open
button. The path and name of
the file displays in the text box. (
See
for more information.)
23
To use a file to supply the Excluded Peaks:
a
Select the
Excluded Peaks File
check box.
b
Touch the
Browse
button next to
Excluded Peaks File
.
c
Select the name of the GEP file that contains the excluded
peaks, and touch the
Open
button. The path and name of
the file displays in the text box. (
See
for more information.
)
24
Touch the
Next
button .
The
opens.
25
For ProteinEXact runs, type the
Total Ladder Concentration
of
the reagent into the text box.
26
Review the selections displayed. To change any selections, use
the
Back
and
Next
buttons to make the desired changes.
27
Touch the
Start
button to start the assay. The
displays the status of the run.
28
After the run has started, you can:
•
Change the exported properties in the Well Table or Peak
Table (see
).
•
Edit the sample names (see
•
Edit the analysis settings (see
).
•
View the raw data from the run (see
)