Optics Normalization Tab
Use the Optics Normalization Tab to calibrate the instrument using
Test Chip C as the calibration standard.To open the Optics
Normalization Tab, touch the
Tools button
on the
and then touch the
Optics Normalization
button.
Figure 33. Optics Normalization Tab
The Optics Normalization Tab contains the following options and
buttons:
Reference Contrast
- The Contrast value from Test Chip C that will
be compared to the scan results Contrast.
Instrument Gain Correction
- The current correction factor for the
instrument.
Chip Temperature (Target 35) degC
- The current temperature of
the chip. The chip is warmed to the target temperature to more
accurately calibrate the optics.
Background
- The background measured on the current instrument
after focusing.
Contrast
- The contrast measured on the current instrument.
New Correction Factor
- The correction factor calculated from the
instrument (Reference Contrast / Scan Results Contrast).