Instrument Safety
10
P/N CLS151164 Rev. BLabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Laser Safety
WARNING
BRIGHT LIGHT HAZARD. LabChip GX /GXII Touch Instruments contain
Class 3B laser diodes. The LabChip GX Touch/GXII Touch are classified as a
Class 1 device because the lasers are appropriately enclosed (embedded)
and indicated with Warning labels.
Complies with 21 CFR 1040.10 except for deviations pursuant to Laser
Notice 50, dated June 24, 2007.
Complies with IEC 60825-1: 1993, A1: 1997, A2: 2001.
635 nm (visible red) laser source, 10 mW maximum continuous
(CW)
WARNING
•
Use of controls or adjustments or performance of procedures
other than those specified herein may result in hazardous
radiation exposure.
•
NEVER remove back, side, or front panels of the instrument
while the laser is powered. Panels (which, if removed, could
lead to laser exposure) are marked with the labels shown below:
•
These panels are intended to be removed for service only by
qualified personnel; they are not intended to be removed during
operation or for maintenance by users. The only removable
maintenance panel is the lower panel at the back of the
instrument, which can be removed to access the back of the
robot, if cleaning is necessary.
Class 1
Laser
Product