Software Reference
121
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Sample Name column
- Displays the sample names, either default
names or names from an open Sample Names file. Sample Names
can be edited until the well read is complete. The Sample Name is
disabled after the well is read.
User Comment
- Displays a comment for the well. User comments
can be edited until the well read is complete. The User Comment is
disabled after the well is read.
Import button
- Opens the selected Sample Names file (*.CSV). If
a run is in progress, touch the Apply button to use the sample
names and user comments for any wells that have not been
completed.
Export button
- Saves the current sample names and user
comments to a Sample Names file (*.CSV) with the specified file
name. The sample names are saved in the file in the same order as
displayed in the table.
Blank 96 Wells button
- Opens a blank Sample Names file with 96
wells.
Blank 384 Wells button
- Opens a blank Sample Names file with
384 wells.
Apply button
- Applies the changes in the table to the currently
running plate. (If a Sample Names file was used to supply the
sample names, the Sample Names file is not updated. Only the
sample names for the run are updated.) Changes on the Sample
Names tab are automatically saved when you touch another tab or
when a well in the currently running plate is completed. If changes
have been made to the currently running well but the changes have
not been applied before the well is complete, the changes to the
current well are ignored, but any changes to the uncompleted wells
are applied.