Troubleshooting and Diagnostics
151
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Current Leakage Check Failed
Warning Message: “Current leakage check failed.”
Chip Temperature Warning
Warning Message: “Average chip temperature is out of range.”
GUI Cannot Connect to CFR Database
Warning Message: “GUI Cannot connect to CFR database. Please
Remap the Server or Contact Administrator.”
Possible Causes
Recommended Actions
Dirty electrode block.
Clean the O-rings (see
Run the Current Leak Diagnostic Test (see
If the problem is not resolved, contact PerkinElmer
Technical Support (see
Possible Causes
Recommended Actions
Chip heater failed.
Restart the instrument. If problem persists, contact
PerkinElmer Technical Support (see
Possible Causes
Recommended Actions
1. Network cable
unplugged or not
connected to the network.
Make sure the network cable is connected to the
LabChip GX Touch (see page
).
Make sure the LabChip GX Touch is logged into the
network.
Make sure the computer where the CDR is installed is
connected to the network.
2. Not logged into the
wireless network.
Make sure the LabChip GX Touch computer is
connected to the network.
3. Software has not been
mapped to the
CDR/Database server.
Open the
Type the Server Name or IP Address of the computer
where the CDR Server software is installed.
If this is the first time connecting to the CDR, see
Cannot Connect to Remote CDR Server on page 164
for server configuration requirements.