Running an Assay
41
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Removing the Plate, Buffer Tube, and Ladder
Tube
To remove the plate, Buffer tube, and Ladder tube from the
instrument when the run is complete:
1
Touch the
Home
button on the
.
WARNING
Pinch Hazard. Keep hands away from plate carrier when carrier is
moving.
2
Touch the
Unload Plate
button on the
. Wait until
the plate carrier is fully extended.
3
Remove the plate from the plate carrier.
4
If necessary, clean the plate carrier with DI water or 70%-
isopropanol solution.
5
Remove the Buffer and Ladder tubes if necessary.
WARNING
Pinch Hazard. Keep hands away from plate carrier when carrier is
moving.
6
Touch the
Load Plate
button on the Home window to retract the
plate carrier.
Removing the Chip
To remove the chip from the instrument:
1
Touch the
Home
button on the
.
2
Touch the
Unload Chip
button on the
. The chip
door on the front of the instrument unlatches.
3
Lift the chip door up.
4
Remove the chip from the chip holder.
5
Close the chip door and press down to latch the door.