Troubleshooting and Diagnostics
155
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
System Components Test
Any chip can be loaded
Chip Interface Test
Remove chip
Description
Potential Cause of Failure
Verify Fan Operation
Damaged fan. Contact PerkinElmer
Technical Support (see
).
Check Chip Cartridge
Interlock
Front door is not open or closed at the
correct time. Close or open front door as
instructed.
Interlock Switch is defective. Contact
PerkinElmer Technical Support (see
Optics Temperature
Optics temperature is not in valid range.
Chassis Temperature1
Chassis temperature is not in valid
range.
RH Temperature
RH temperature is not in valid range.
Description
Potential Cause of Failure
Marker Pressure
Sensor Test
Pressure sensor out of calibration.
Contact PerkinElmer Technical Support
(see
Marker Pressure Chip
Interface Clog Test
Clog in pressure line. Contact
PerkinElmer Technical Support (see
Priming Pressure
Sensor Test
Pressure sensor out of calibration.
Contact PerkinElmer Technical Support
(see
Priming Pressure Chip
Interface Clog Test
Clog in pressure line. Contact
PerkinElmer Technical Support (see
Sipping Pressure
Sensor Test
Pressure sensor out of calibration.
Contact PerkinElmer Technical Support
(see
Sipping Pressure Chip
Interface Clog Test
Clog in pressure line. Contact
PerkinElmer Technical Support (see