Software Reference
80
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
Plate Diagram
- Select the wells to read during the run. Selected
wells are blue, unselected wells are white.
•
To select all rows on the plate, touch the
Select All Rows
button at the top left corner of the plate diagram.
•
To select a row, touch the row letter on the left side of the plate.
•
To select a column, touch the column number at the top of the
plate.
•
Clear a single well by touching the selected well.
Full Size button
- Enlarges the Plate Diagram to full screen to
enable easier selection of wells, especially when selecting wells on
384-well plates. Touch the
Done
button in the upper right corner to
return to the Select Wells tab.
Quadrants
- Touch the
1
,
2
,
3
, or
4
buttons to select or clear all
wells in each quadrant. Any wells that are already selected are not
changed unless the wells are in the specified quadrant.
Column-wise check box
- Specifies the order in which the wells
are sampled during the assay:
•
If selected, the wells in each column are sampled from top to
bottom before proceeding to the next column (A1, B1, C1... A2,
B2, C2...). Ladders are sipped every 12 wells, so some columns
will include a ladder in the column.
•
If cleared, the wells in each row are sampled from left to right
before proceeding to the next row (A1, A2, A3... B1, B2, B3...).
Ladders are sipped every 12 wells.
Selection Sip Order check box
- If selected, the wells are sipped
in the order in which the wells are selected. Selecting a rectangular
region orders the wells from the start of the selection region to the
end of the selection region, either row-wise or column-wise. If not
selected, the wells are sipped in row or column order.
Advanced Settings
- If selected, uses the Advanced Run Settings:
•
Sample Sips
- Specifies the number of times to run the selected
wells. Each well is sampled once, and then the entire run
repeats. The data from all wells is combined in one data file.
•
Plate Cycles
- Specifies how many times to repeat the assay on
the plate.
•
Random Select%
- Specifies the percent of wells to randomly
sample from the selected wells. For example, if 10 wells are
selected, and the Random Select% is set to 50%, five wells,
selected at random, will be sampled during the run.
Next button
- Opens the
.