Excluded Peak File
- If selected, specifies the file that will supply
the Excluded Peaks for the data file. Touch the
Browse
button next
to Excluded Peaks File, select the name of the GEP file that
contains the excluded peaks, and touch the Open button. The path
and name of the file displays in the text box. Use the LabChip GX
Reviewer software to create the Excluded Peak file. See
for more information about Excluded Peak files.
The Setup Run tab contains the following buttons:
•
Back
- Opens the
to change the selected wells
or sampling options.
•
Next
- Opens the
.
Start Run Tab
Use the Start Run tab to review the settings for the run, and start
the run if the settings are correct. To open the Start Run tab, touch
the
Next
button on the
Start Run
button on top of the
.
Figure 25. Start Run Tab
The Start Run Tab contains:
Chip Status
- Displays the Assay Type, Chip Reagent Expiration,
Chip Expiration date, and Chip Life.