Washing the Chip
Some protein samples may have components which produce data
with extra peaks, spikes or other artifacts. When these artifacts are
present, washing protein chips on the LabChip GXII Touch
immediately before the next use can often restore data quality.
1
Prepare the chip and buffer tube as instructed in the
Assay User
Guide
.
2
Insert the prepared chip and buffer tube into the instrument (see
3
Touch the
Home
button on the
.
4
Touch the
Wash
button on the
.
5
When the wash is complete, follow the instructions in the
Assay
User Guide
to run the chip.