Software Reference
85
P/N CLS151164 Rev. B
LabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
•
Type
- Displays the type of chip in the instrument, or the assay
type selected for the chip if the chip supports multiple assays.
•
Chip Reagents
- Displays the time and number of samples
(whichever comes first) until the chip reagents expire. Reagent
expiration depends on the chip type. The expiration time and
number of samples reset whenever the front door is opened.
•
Chip Expiration
- The expiration date of the chip. The
expiration date is stored on the RF ID tag on the chip.
•
Chip Life
- Displays the number of samples until the chip
expires. Chip Life depends on the chip type.
Run Parameters
- Displays the settings selected on the
and the
.
Plate Diagram
- The wells selected to read are blue. Wells that are
not selected to read are white.
Total Ladder Concentration text box
- (ProteinEXact assays only)
Type the total ladder concentration of the reagent into the text box.
Back button
- Opens the
Export button
- Opens the
Save Run Parameters to File
window to export the current run settings to a run file (.xml).
Start button
- Runs the plate using the current settings.
The
opens to view the status of the run or priming
and calibration.