Instrument Safety
12
P/N CLS151164 Rev. BLabChip GX Touch/GXII Touch User Manual
PerkinElmer, Inc.
International
All power cord sets must be approved by an acceptable, accredited
agency responsible for evaluation in the country where the power
cord set and system will be used.
The flexible cord must be <HAR> Type H05VV-F, 3-conductor,
minimum 0.75 - 1 mm
2
conductor size (230 volt input). Power cord
set fittings that is, the appliance coupler and wall plug, must bear
the certification mark of the agency responsible for evaluation in the
country where it will be used. The appliance coupler must meet the
mechanical configuration of an EN 60 320/IEC 320 Standard Sheet
C13 connector for mating with appliance inlet on the system.
Fuses
The LabChip GX Touch/GXII Touch instruments contain two fuses.
See
Changing the Fuses on page 142
instructions. Contact PerkinElmer Technical Support (see
) to
order replacement fuses.