pending triggered level is a stored current value that is transferred to the power supply output
when a trigger occurs. A pending triggered level is unaffected by subsequent CURR commands
and remains in effect until the trigger subsystem receives a trigger or is aborted with an ABORt
command. In order for CURR:TRIG to be executed, the trigger subsystem must be initiated (see
INITiate).
CURR? and CURR:TRIG? return presently programmed immediate and triggered levels. If no
triggered level is programmed, then the CURR level is returned for both values. CURR? MAX
and CURR? MIN return the maximum and minimum programmable immediate current levels.
CURR:TRIG? MAX and CURR:TRIG? MIN return the maximum and minimum programmable
triggered current levels.
Command Syntax:
[SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude] <NRf+>
[SOURce]:CURRent[:LEVel]:TRIGgered [:AMPLitude] <NRf+>
Examples: CURR
200
CURRENT:LEVEL MIN
CURRENT:LEVEL:IMMEDIATE:AMPLITUDE 2.5
CURR:TRIG 20
CURRENT:LEVEL:TRIGGERED 20
Query Syntax:
[SOURce]:CURRent[:LEVel][:IMMediate][:AMPLitude]?
[SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude]? MAX
[SOURce]:CURRent[:LEVel] [:IMMediate][:AMPLitude]? MIN
[SOURce]:CURRent[:LEVel]:TRIGgered [:AMPLitude]?
[SOURce]:CURRent[:LEVel]:TRIGgered [:AMPLitude]? MAX
[SOURce]:CURRent[:LEVel]:TRIGgered [:AMPLitude]? MIN
Returned Parameter: <NR2>
Related Commands (for CURR):
*SAV, *RCL, *RST
Related Commands (for CURR:TRIG):
ABOR, CURR, *RST
4.3.1.4 CURR:PROT
This command sets the over current trip (OCT) level of the power supply. If the output current
exceeds the OCT level, then the power supply output is disabled and the Questionable Condition
status register OC bit is set (see Table 4.2). An over current trip condition can be cleared with the
OUTP:PROT:CLE command after the condition that caused the OCT trip is removed.
CURR:PROT? returns presently programmed OCT level. CURR:PROT? MAX and
CURR:PROT? MIN return the maximum and minimum programmable OCP levels.
Command Syntax:
[SOURce]:CURRent:PROTection[:LEVel] <NRf+>
52
Summary of Contents for XR III series
Page 1: ...OPERATING AND SERVICE MANUAL XR SERIES III DC POWER SUPPLIES...
Page 2: ......
Page 3: ...MAGNA POWER ELECTRONICS INC 39 ROYAL ROAD FLEMINGTON NJ 08822 February 20 2012...
Page 4: ......
Page 88: ...Figure 4 1 Status Byte Generation Figure 4 2 ESE and ESR Generation 76...
Page 95: ...IEEE Standard CLS ESR ESE STB SRE IDN SAV RCL RST Notes 1 C command Q query 83...
Page 97: ...Figure 5 1 Configuration setup Figure 5 2 GPIB communications setup 85...
Page 99: ...Figure 5 4 Virtual Control Panel Figure 5 5 Command Panel 87...
Page 102: ...Figure 5 7 Calibration Panel Figure 5 8 Firmware Panel 90...
Page 103: ...Figure 5 9 Modulation Panel 91...
Page 123: ...Figure B 1 Information Panel Figure B 2 Configure Panel 111...
Page 124: ...Figure B 3 Reboot in Progress Panel Figure B 4 Web Control Panel 112...