MEP 660 - Trouble shooting Manual
Doc No.: IM – 51
Rev.2.0
Print: October 2008
144
slack (cable ID: AY). Tighten the connector and check the
voltage again.
2) If voltage is still zero, check the continuity of BKBCL wire
from the magnet valve to MEP-660 (Cable ID: Pin No.L). If
necessary replace the cable and check the voltage once
again.
3) If voltage is still zero, then the digital output card MLSD 16
in slot No.14 defective. Replace the card.
8. Repeat the test again vide item No.6 and ensure BKT is throwing to
braking side