SERVICE MANUAL
5–90
10FEB97 – SM3226-1
PH4 R Magazine Door Sensor
Note
This procedure cannot be done on machines after S/N 500249. DETECTORS were changed from
PHOTODETECTORS to MICROSWITCHES to comply with European standards.
To Check:
[1] Install the PORTABLE COMPUTER. See “Using the Portable Computer” in the Diagnostic Section.
[2] Remove the RIGHT IMAGE UNIT COVER. If necessary, see the procedure on Page 2–3.
[3] Use the PORTABLE COMPUTER to display the
CES Main Menu.
[4] Enter [1] for “Monitor Mode”.
[5] Enter [5] for “Sensors”.
[6] On the PORTABLE COMPUTER, check that the R MAGAZINE DOOR SENSOR PH4 reads 0 when the
DOOR is closed, and 1 when the DOOR is open.
To Adjust:
[7] Loosen the 2 SCREWS.
[8] Move the SENSOR BRACKET until the SENSOR reads 0 when the DOOR is closed and 1 when the DOOR
is open.
[9] Tighten the 2 SCREWS.
MAIN MENU
1 Monitor Mode 4 System Params
2 System Logs 5 Debug Utility
3 Boards Diag 6 Exit
Enter Menu Item:
MONITOR MODE MENU
1 Motors 6 Seq Board Diag
2 Solenoids 7 Enter Commands
3 Relay 8 Set Beam-Splitter
4 Miscellaneous 9 Auto Sensor Test
5 Sensors 10 Semi-Auto Sensor Test
Enter Menu Item:
H129_1908BC
H129_1908BCC
SCREW (2)
SENSOR
BRACKET
R MAGAZINE
DOOR SENSOR