SERVICE MANUAL
5–32
10FEB97 – SM3226-1
[5] Observe the wavelength on the OSCILLOSCOPE. The pulse width should be between 20 ms and 22 ms.
To Adjsut:
[6] Rotate the correct POTENTIOMETER for each sensor until the value is within specification.
PHT2: VR10 for the SLOW SCAN FILM SENSOR
PHT3: VR11 for the FILM PROCESSOR ADAPTER FILM SENSOR
PHT1: VR12 for the TRANSPORT GUIDE PLATE JAM SENSOR
[7] Connect J218 to the SEQUENCE BOARD.
[8] Remove TL-4715.
20 ms.
Approx.
5 V
GND
Approx. 30 ms.
H129_1911BC