SERVICE MANUAL
4–10
10FEB97 – SM3226-1
[8] Press the left-side [CHAN] key under the “P1”
three times when the DENSITOMETER displays
the screen at the right, with a “P1” at the bottom
left corner.
Note
The third time the [CHAN] key is pressed, the
DENSITOMETER will display:
Important
The DENSITOMETER is ready to display the densities
for all 21 steps that it has read from the Calibration
Image. The Kodak Ektascan 1120 LASER PRINTER
uses only the densities of steps 7 through 15. To read
and record the required densities:
[9] Press the right-side [CHAN] key and scroll
through the densities.
[10] Record the values when of density steps 7 through
15.
[11] Check the values by scrolling through density
steps 7 through 15 again.
[12] Simultaneously press [MENU] and [MENU].
Note
The PORTABLE COMPUTER will display the Main
Menu.
[13] Enter the values as described in the procedure
“Using a Stand-alone DENSITOMETER”
beginning in Step 4 on Page 4–6.
CH:I SD.Indx = x.xx
P1 CIndx = x.xx
CH:I Temp F = xx.x
P4 Den #01 = x.xx