Operation Adjustments
SM3226-1 – 10FEB97
4–7
[8] Check the density values by pressing [NEXT] and
reviewing each of the 9 density values.
[9] Press [SELECT].
[10] Press [SELECT] again to begin the calibration.
Note
• The PRINTER will compute the calibration table. Calibration will take about one minute. When the calibration
has been calculated successfully, the LASER PRINTER displays:
[ 00 PRINTER READY ] .
• If calibration is not successful, the CONTROL PANEL will display error 54. See the Calibration Errors
procedure below.
Calibration Errors
If calibration is unsuccessful, the LASER PRINTER displays an error message.
[1] Check that the LASER PRINTER and the processor are warmed up completely.
1D BEGIN CALIBRATION
CALIBRATING
54 ERROR:
CALIBRATION