Diagnostic Check Procedures
SM3226-1 – 10FEB97
3–13
[26] Enter [R] [E] [C] to move the film to the
RECEIVE MAGAZINE.
[27] Enter the correct number corresponding to the size
of the film loaded into the SUPPLY MAGAZINE.
• 0: 8x10
• 1: 11x14
• 2: 35x35
• 3: 35x43
[28] Press [ESC] 3 times to exit diagnostics.
[29] Remove the VACUUM GAUGE.
[30] Install the CAP.
[31] Install the RIGHT IMAGE UNIT COVER.
error=0
Enter command: REC
Film size? (0-3):
Film size? (0-3):X
H129_0103DC
H129_0103DCA
CAP
GAUGE
VACUUM