SERVICE MANUAL
5–26
10FEB97 – SM3226-1
Important
If necessary, do the adjustment procedure in the following step.
DRIVER BOARD
To Adjust:
[7] Rotate the correct POTENTIOMETER for each sensor until the value is within specification.
IL1: VR3 for the SLOW SCAN FILM SENSOR
IL2: VR4 for the FILM PROCESSOR ADAPTER FILM SENSOR
IL3: VR13 for the TRANSPORT GUIDE PLATE JAM SENSOR
[8] Replace connectors J211, J212, and J213.
23
24
21
22
2
1
H129_1724DC
VR4
P211
VR11
VR10
VR12
P213
VR3
VR13
P212
CP5
CPG5
U14
PHT2
PHT1
PHT3
H129_1724DCB
VR6
VR5