[11] Enter [I] [T] [R] to place the TRANSPORT
ROLLERS in the initial position.
[12] Enter [I] [F] [D] to move the SUCKER PADS into
the initial position.
[13] Check that the distance between the SLIDE BASE and the ROLLER is 1.5 mm (0.06 in.).
Enter Command: ITR
error=0
Enter Command: IFD
Check the S-Mag open.
Are you ready?,Y/N (0/1) [0]:
error=0
Enter Command:
H129_5404DC
H129_5404DCA
1.5 mm
BASE
SLIDE
ROLLER