SERVICE MANUAL
4–22
10FEB97 – SM3226-1
CES Hidden Mode Setup Procedure
Use this procedure to enter parameters close to those obtained during calibration whenever:
• installing a new BATTERY
• installing a new SEQUENCE BOARD
• installing a new software version
[1] Press the [
▼
] key, the [
▲
] key, and the POWER
SWITCH to 1 at the same time.
[2] Release the switches, and then press the ACCESS
SWITCH within 5 seconds.
Note
The CONTROL PANEL will display messages as they
are shown in the table on the opposite page.
[3] Check the ALIGNMENT VALUE LABEL
located in front of the CARD RACK to determine
a value for the parameter being entered or
changed.
[4] Use the [NEXT] key to step through the options
displayed ont he CONTROL PANEL while
entering the parameters required.
[5] When entering values, do the following:
• Press the [
▼
] and [
▲
] keys to select a value.
• Press [NEXT] to advance to the next
parameter when the correct value is
displayed.
• Press [SELECT] to enter changed values and
return to normal operating mode.
H129_0709AC
VALUE LABEL
ALIGNMENT
ACCESS
SWITCH
POWER SWITCH
H129_0709ACB
H129_2500AC
READY
PRINTER
00
NEXT
SWITCH
SELECT
SWITCH
DOWN
SWITCH
UP
SWITCH
H129_2500ACA