SERVICE MANUAL
5–8
10FEB97 – SM3226-1
Important
If necessary, do the adjustment procedure in the following step.
DRIVER BOARD
To Adjust:
[7] Rotate the correct POTENTIOMETER for each sensor until the value is within specification.
IL1: VR3 for the SLOW SCAN FILM SENSOR
IL2: VR4 for the FILM PROCESSOR ADAPTER FILM SENSOR
IL3: VR13 for the TRANSPORT GUIDE PLATE JAM SENSOR
[8] Replace connectors J211, J212, and J213.
29
30
39
40
1
2
1
2
H129_1706DC
1
2
22
21 1
2
23
24
CNA
CNB
CP5
POTENTIOMETER
VR4
P211
P213
P212
POTENTIOMETER
VR3
POTENTIOMETER
VR13
H129_1706DCA