SERVICE MANUAL
5–76
10FEB97 – SM3226-1
PHS1 S Magazine Open Sensor
To Check:
[1] Install the PORTABLE COMPUTER. See
“Using the Portable Computer” in the Diagnostic
Section.
[2] Use the PORTABLE COMPUTER to display the
CES Main Menu.
[3] Enter [1] for “Monitor Mode”.
[4] Enter [7] for “Enter Commands”.
Note
• Commands must be entered using capital letters.
• See the Diagnostics section to clear any error
messages displayed on the PORTABLE
COMPUTER during this procedure.
[5] Open the RIGHT IMAGE UNIT COVER. If
necessary, see the procedure on Page 2–3.
[6] Enter [S] [M] [O] to open the SUPPLY
MAGAZINE WINDOW.
MAIN MENU
1 Monitor Mode 4 System Params
2 System Logs 5 Debug Utility
3 Boards Diag 6 Exit
Enter Menu Item:
MONITOR MODE MENU
1 Motors 6 Seq Board Diag
2 Solenoids 7 Enter Commands
3 Relay 8 Set Beam-Splitter
4 Miscellaneous 9 Auto Sensor Test
5 Sensors 10 Semi-Auto Sensor Test
Enter Menu Item:
Input ‘HLP’ to know commands.
Enter command:
Enter Command: SMO
error=0