Operation Adjustments
SM3226-1 – 10FEB97
4–11
Checking Stored Calibration Settings
[1] Press [ON-LINE].
[2] Press [
▼
] and [
▲
] keys to select the desired
calibration number.
[3] Press [SELECT].
[4] Press [SELECT].
[5] Check that the stored density is underscored.
Note
If the stored density is not underscored, a new
calibration must be done. See Calibrating the Printer on
Page 4–4.
[6] Press [SELECT].
[7] Press [NEXT].
[8] Press [SELECT].
[9] Wait one minute for the PRINTER to compute the
calibration table.
[10] Press and release [NEXT] to check each of the
density values.
[11] Press [SELECT].
[12] Press [SELECT].
1. CALIBRATION #: X
1A SET DENSITY
-3 -2 -1 0 +1 +2 +3
1B PRINTING CAL IMAGE
1C ENTER DATA
1. CALIBRATION #: X
ENTER STEP 1: X
ENTER STEP 9: X
1D BEGIN CALIBRATION
CALIBRATING
00 PRINTER READY