Electrical Adjustments
SM3226-1 – 10FEB97
5–95
[20] Check that the LED indicator on the underside of
the SUPPLY ROLLER POSITION SENSOR is
off.
Note
If the LED indicator is off after the previous step,
STOP, this procedure has been completed.
[21] Press the [ESC] key on the PORTABLE
COMPUTER 2 times to return to the MAIN
MENU.
[22] Manually, align the SUPPLY ROLLER shaft so
that the SUPPLY ROLLER rests against the
STOP BRACKET and TRANSPORT
ROLLERS. The ROLLER will remain in place.
[23] Enter [1] for “Monitor Mode”.
[24] Enter [7] for “Enter Commands”.
MONITOR MODE MENU
1 Motors 6 Seq Board Diag
2 Solenoids 7 Enter Commands
3 Relay 8 Set Beam-Splitter
4 Miscellaneous 9 Auto Sensor Test
5 Sensors 10 Semi-Auto Sensor Test
Enter Menu Item:
Input ‘HLP’ to know commands.
Enter command:
H129_4201CC
FLAG
SENSOR
POSITION
ROLLER
SUPPLY
ROLLER
TRANSPORT
ROLLER
SUPPLY
BRACKET
STOP
H129_4201CCB