SERVICE MANUAL
5–94
10FEB97 – SM3226-1
[7] Enter [7] for “Enter Commands”.
Note
• Commands must be entered using capital letters.
• See the Diagnostics section to clear any error
messages displayed on the PORTABLE
COMPUTER during this procedure.
[8] Enter [I] [T] [R] to initialize the roller.
[9] Enter [P] [M] [S].
[10] Enter [1] [4] for the pulse number.
[11] Enter [1] [0] for the speed.
[12] Check that the SUPPLY ROLLER is resting
against the STOP BRACKET and in contact with
the TRANSPORT ROLLERS. If the SUPPLY
ROLLER is not in the correct position, go to Step
21.
Note
The following steps allow the SUPPLY ROLLER
STEPPER MOTOR to be rotated in incremental steps
equal to the number of pulses entered. The number of
pulses entered will depend on how far away the edge of
the flag is from the sensor. The flag should activate the
sensor within 10 pulses after the initial -64 pulses are
entered.
[13] Enter [P] [M] [S] to start the SUPPLY ROLLER
MOTOR.
[14] Enter [-] [6] [4] for the number of pulses.
[15] Enter [3] [0] [0] for the speed.
[16] Check that the LED indicator on the underside of
the SUPPLY ROLLER POSITION SENSOR is
on. If the LED indicator is off, go to Step 21.
[17] Enter [P] [M] [S] to start the SUPPLY ROLLER
MOTOR.
[18] Enter [-] [1] [0] for the number of pulses.
[19] Enter [1] [0] for the speed.
Input ‘HLP’ to know commands.
Enter command:
Enter Command: ITR
error=0
error=0
Enter Command: PMS
Pulse number? (-30000 to 30000)[0]: 14
Speed? (10 to 1500 pps) [10]: 10
Enter Command: PMS
How many pulses? (-30000 - 30000):
How many pulses? (-30000 - 30000): -64
Speed? (10 - 1500 pps):
Speed? (10 - 1500 pps): 300
error=0
Enter command:
Enter Command: PMS
How many pulses? (-30000 - 30000):
How many pulses? (-30000 - 30000): -10
Speed? (10 - 1500 pps):
Speed? (10 - 1500 pps): 10
error=0
Enter command: