Operation Adjustments
SM3226-1 – 10FEB97
4–23
The following is a list of parameters in the sequence in which they are displayed:
Parameter
Description
Enter this value:
“BS PULSE xx:yyy:
To set the beam splitter pulses from 0 to
200.
See LABEL for values.
“FILM SIZE: xxxxx”
To set the film size.
35x43/35x35/14x11/8x10
“FILM OUTPUT: xxxxx”
To set the film output.
Proc/R-Mag
“MAX RECEIVE: xxx”
To set the maximum number of film
sheets for the Receive Magazine.
80/100 *
“R-MAG FILM: xxx”
To read the number of film sheets in the
Receive Magazine.
0
“FST: xx.xxmm”
To set the film start position.
See LABEL for values.
“IMST (8x10): xx.xxmm”
To set the 8x10 image start position.
“IMST: xx.xxmm”
To set an image start position other than
8x10.
“PEST: xxx.xxmm”
To set the Polygon edge start position.
“PEEND: xxx.xxmm”
To set the Polygon edge end position.
“SLOW: xx.xxmm”
To set the Slow Scan direction position.
“PRINT COUNT: xxx”
To set the number of test prints.
1
“JAM ERROR: xx”
To store the jam error.
0
“ACTIVE CAL: X”
1-11
“DENSITY: xx”
To set the density. (-3 to +3)
0
“RAW DATA x:y_yy”
To enter data for calibration
(1)0.15, (2)0.22, (3)0.42,
(4)0.79, (5)1.25, (6)1.73,
(7)2.12, (8)2.46, (9)2.71 **
“CAL DATA: xxxx”
To use the previously entered raw data to
calibrate the PRINTER when calibration
error occurs. See the procedure on Page
4–12 for more information.
OK
“CONTRAST: xx”
To set the contrast for test prints.
0
“LOW STEP: xxx”
To set the number of pulses for the “Film
Supply Low” message. (162 - 222)
Enter desired value, See
label, default=192
“USER CAL: xxxxxxx”
To set the permission/inhibition for
calibration.
PERMIT
“TRANSPORT:
NORMAL/SPECIAL”
To correct vertical banding.
NORMAL ***
“PROC TYPE:”
To choose the PROCESSOR that is
docked to the PRINTER.
Choose “M35/270RA” or
“M7B”. ****
*
If using 8 x 10 in. FILM, select “80” for the maximum receive value. If using 35 x 43 cm.
select either “80” or “100”, depending on the particular needs of the customer.
**
Do not enter new values for Raw Data after Calibration has been completed.
***
Changing “TRANSPORT” from “NORMAL” to “SPECIAL” will correct vertical banding
artifacts.
**** Used only on new PRINTERS, or old PRINTERS that have had Modification 6 installed.