[26] Press {ESC] to return to the Beam-splitter menu.
[27] Enter [5] to set each pulse.
[28] Enter each of the new pulse values recorded in
Steps 24 and 25.
[29] Press [ESC] 4 times to exit diagnostics.
SET BEAM-SPLITTER MENU
1 Print Grey Scale 4 Show Pulse
2 Enter Density 5 Set Each Pulse
3 Set Normal Status
Enter Menu Item: 5
Density Current New
-3 xxx :
-2 xxx :
-1 xxx :
0 xxx :
+1 xxx :
+2 xxx :
+3 xxx :