SERVICE MANUAL
3–6
10FEB97 – SM3226-1
[8] Enter [1] for “Monitor Mode”.
[9] Enter [7] for “Enter Commands”.
Note
• Commands must be entered using capital letters.
• See the Diagnostics section to clear any error
messages displayed on the PORTABLE
COMPUTER during this procedure.
Style A
[10] For the STYLE A BOARD, enter [L] [D] [O] to
energize the laser.
For the STYLE B BOARD, enter [O] [L] [T], [1], and
[4095].
Style B
Important
The BEAM SPLITTER must be reset to the original alignment value to measure laser power. If the alignment value
was changed to meet customer needs, that value must be re-entered after the laser power measurement has been
taken.
[11] Enter [C] [B] [S] to set the BEAM SPLITTER
density settings.
[12] Enter [0] for density setting 0.
[13] Enter [S] [E] [T].
[14] Enter [0].
MONITOR MODE MENU
1 Motors 5 Sensors
2 Solenoids 6 Board Diag
3 Relay 7 Enter Commands
4 Miscellaneous 8 Set Beam-splitter
Enter Menu Item:
Input ‘HLP’ to know commands.
Enter command:
Enter Command: LDO
Enter Command: OLT
Linear or constant (0/1)[0]:1
Look up table data (0 to 4095)[0]:4095
error=0
Enter command:
Enter Command: CBS
Density? (-3 to 3)[-3]:
Enter Command: CBS
Density? (-3 to 3)[-3]:0
Current Density: 0
error=0
Enter Command: SET
Density? (-3 to 3)[-3]:0
Old pulse number: XXX
Pulse number[B]? (0 to 200):XXX