186
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
12
Acquisition Control
Using Peak Detect Mode to Find a Glitch
1
Connect a signal to the oscilloscope and obtain a stable display.
2
To find the glitch, press the
[Acquire]
key; then, press the
Acq Mode
softkey until
Peak Detect
is selected.
3
Press the
[Display]
key then press the
∞
Persistence
(infinite persistence) softkey.
Infinite persistence updates the display with new acquisitions but does not
erase previous acquisitions. New sample points are shown at normal intensity
while previous acquisitions are displayed at reduced intensity. Waveform
persistence is not kept beyond the display area boundary.
Press the
Clear Display
softkey to erase previously acquired points. The display
will accumulate points until
∞
Persistence
is turned off.
4
Characterize the glitch with Zoom mode:
a
Press the
zoom key (or press the
[Horiz]
key and then the
Zoom
softkey).
b
To obtain a better resolution of the glitch, expand the time base.
Use the horizontal position knob (
) to pan through the waveform to set the
expanded portion of the normal window around the glitch.
Averaging Acquisition Mode
The Averaging mode lets you average multiple acquisitions together to reduce
noise and increase vertical resolution (at all time/div settings). Averaging requires
a stable trigger.
The number of averages can be set from 2 to 65536 in power-of-2 increments.
A higher number of averages reduces noise more and increases vertical resolution.
# Avgs
Bits of resolution
2
8
4
9
16
10
64
11
≥
256
12
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...