Math Waveforms
4
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
79
FFT Measurement
FFT is used to compute the fast Fourier transform using analog input channels or
an arithmetic operation g(t). FFT takes the digitized time record of the specified
source and transforms it to the frequency domain. When the FFT function is
selected, the FFT spectrum is plotted on the oscilloscope display as magnitude in
dBV versus frequency. The readout for the horizontal axis changes from time to
frequency (Hertz) and the vertical readout changes from volts to dB.
Use the FFT function to find crosstalk problems, to find distortion problems in
analog waveforms caused by amplifier non-linearity, or for adjusting analog filters.
To display a FFT waveform:
1
Press the
[Math]
key, press the
Function
softkey and select
f(t)
, press the
Operator
softkey and select
FFT
.
•
Source 1
— selects the source for the FFT. (See
filters on an arithmetic operation"
on page 71 for information about using
g(t)
as the source.)
•
Span
— sets the overall width of the FFT spectrum that you see on the
display (left to right). Divide span by 10 to calculate the number of Hertz per
division. It is possible to set Span above the maximum available frequency, in
which case the displayed spectrum will not take up the whole screen. Press
the
Span
softkey, then turn the Entry knob to set the desired frequency span
of the display.
•
Center
— sets the FFT spectrum frequency represented at the center vertical
grid line of the display. It is possible to set the Center to values below half
the span or above the maximum available frequency, in which case the
displayed spectrum will not take up the whole screen. Press the
Center
softkey, then turn the Entry knob to set the desired center frequency of the
display.
•
Scale
— lets you set your own vertical scale factors for FFT expressed in
dB/div (decibels/division). See
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...