CAN/LIN Triggering and Serial Decode
23
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
337
For LIN 1.2 signals, use the LIN 1.3 setting. The LIN 1.3 setting assumes the
signal follows the "Table of Valid ID Values" as shown in section A.2 of the LIN
Specification dated December 12, 2002. If your signal does not comply with the
table, use the LIN 2.0 setting.
12
Press the
Sync Break
softkey and select the minimum number of clocks that
define a sync break in your LIN signal.
LIN Triggering
To set up the oscilloscope to capture a LIN signal, see
LIN triggering can trigger on the rising edge at the Sync Break exit of the LIN
single-wire bus signal (that marks the beginning of the message frame), the
Frame ID, or the Frame ID and Data.
A LIN signal message frame is shown below:
1
Press
[Trigger]
.
2
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob to
select the serial slot (Serial 1) on which the CAN signal is being decoded.
3
Press the
Trigger:
softkey; then, turn the Entry knob to select the trigger
condition:
•
Sync
(Sync Break) — The oscilloscope triggers on the rising edge at the Sync
Break exit of the LIN single-wire bus signal that marks the beginning the
message frame.
Sync
Break
Identifier
Break
Sync Break Exit
Sync
Field
Checksum
Field
Data
Fields
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...