Triggers
10
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
167
•
Exit Sus
(Exit Suspend) — triggers when exiting an idle state > 10 ms. This is
used to see the suspend/resume transition.
7
Press the
Speed
softkey to select the speed of the transaction being probed.
You may select Low Speed (1.5 Mb/s) or Full Speed (12 Mb/s).
8
Press the
D+
and
D–
softkeys to select the channel connected to the USB signal
D+ and D– lines. The D+ and D– labels for the source channels are
automatically set.
As you press the
D+
or
D–
softkey (or rotate the Entry knob), the D+ and D– label
for the source channel is automatically set and the channel you select is shown
in the upper-right corner of the display next to "USB".
If you have connected the oscilloscope's analog source channels to the D+ and
D- signals: Adjust the trigger level for each connected analog channel to the
middle of the waveform by pressing the
D+
or
D-
softkey, then turning the
Trigger Level knob.
If you have connected the oscilloscope's digital source channels to the D+ and
D- signals (this applies to MSO model oscilloscopes only): Press the
[Digital]
key
and select
Thresholds
to set an appropriate threshold level for the digital
channels.
The value of the trigger level or digital threshold is displayed in the upper-right
corner of the display.
Serial Trigger
With serial decode option licenses (see
you can enable serial trigger types. To set up these triggers, see:
•
•
•
•
•
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...